banner

Custom Reach-in Chamber

Delivery term:The date of payment from buyers deliver within days
  • Price:

    Negotiable

  • minimum:

  • Total supply:

  • Delivery term:

    The date of payment from buyers deliver within days

  • seat:

    Beijing

  • Validity to:

    Long-term effective

  • Last update:

    2022-06-19 16:42

  • Browse the number:

    123

Send an inquiries
Company Profile
Hunan Climate Instrument Co., Ltd
By certification [File Integrity]
Contact:

kelai(Mr.)  

Email:

Telephone:

Phone:

Area:

Beijing

Address:

Building B5, No. 29, Qixin Road, Dongyong Town, Nansha District, Guangzhou City, Guangdong Province, China

Website: http://www.climate-test.com/ http://kelai.sdjiayuan.net/
Product details

Overview

Glai Custom Reach-in Temperature and Humidity Test Chamber is used to test the performance of materials in various environments and test the heat resistance, cold resistance, dry resistance and humidity resistance of various materials. Suitable for testing the quality of products such as electronics, electrical appliances, mobile phones, communications, meters, vehicles, plastic products, metals, food, chemicals, building materials, medical care, aerospace, etc..


Test Chamber Applications

1. To simulate test environment with different temperature and humidity

2. Cyclic test includes climatic conditions: holding test, cooling-off test, heating-up test, moistening test and drying test…

3. Cable port with flexible silicone plug for cable routing to provide the condition of test unit under operation

4. Uncover test unit weakness in a short-term test with accelerated time effect


Fulfill the Standard:
The performance pointer meets the requirements of GB5170, 2, 3, 5, 6-95, "Basic parameter verification methods for environmental test equipment for electric and electronic products, low temperature, high temperature, constant humidity and heat, alternating humidity and heat test equipment"
Basic environmental test procedures for electric and electronic products Test A: Low temperature test method GB 2423.1-89 (IEC68-2-1)
Basic environmental test procedures for electric and electronic products Test B: High temperature test method GB 2423.2-89 (IEC68-2-2)
Basic environmental test procedures for electric and electronic products Test Ca: Constant humidity test method GB / T 2423.3-93 (IEC68-2-3)
Basic environmental test procedures for electric and electronic products Test Da: Alternating damp heat test method GB / T423.4-93 (IEC68-2-30)

Control methods and features:
The balanced temperature control system (BTHC) controls the SSR in the P.I.D. mode, so that the heating and humidification amount of the system is equal to the damp heat loss amount, so it can be used stably for a long time.


Specification

Sizes of Temperature Chambers

Model

Volume (L)

Temperature Range

(°C)

Rated Power

Internal Dimension (mm) W*H*D

External Dimension (mm) W*H*D

T280

280

-20~150°C

6.1KW

700*800*500

900*2080*1600

-40~150°C

6.1KW

-70~150°C

7.3KW

T500

500

-20~150°C

7.6KW

700*800*900

900*2080*2000

-40~150°C

8.8KW

-70~150°C

8.8KW

T1000

1000

-20~150°C

13.3KW

1000*1000*1000

1200*2280*2100

-40~150°C

15.3KW

-70~150°C

15.5KW

Basic Specification

Performance

Temperature Range

A: -20~150°C, B: -40~150°C, C: -70~150°C, D : or customized

Temperature Fluctuation

±0.5°C

Temperature Uniformity

±2.0°C

Ramp Rate

1~3°C/min for heating (non-linear, based on on-load)

0.7~1.0°C/min for cooling (non-linear, based on on-load)

Note: can be customized as per requirement

Material

External Casing

Cold-rolled A3 (Q235) steel plates with plastics painted.

Interior Working Room

Stainless Steel Plates SUS304 with miror surface

Heat Isolation

Polyurethane foams and fine glass fibres

Illumination Lamp

One illumination lamp in per standard test chamber.

Casters

4pcs, with PU covered for protection. Height adjustable and direction universal.

Viewing Window

Multi-layer hollow toughened glass with conductive films for heating to defrost

Door

Manually operated sliding door with unlock button

Heating & Cooling System

Heater

Steipped Nickel-chromium Wire Heater

Air Blower

Anti-temperature stainless fan blades with low noise

Compressor

Hermetically-sealed Techmesh® or semi-sealed Bitzer® or GEA® or equivalent compressors

Cooling Method

Single or dual cooling (Air-cooled or Water-cooled)

Refrigerant

R404A or/and R23 (eco-friendly)

Condenser

Plate-type heat exchanger

Control System

Controller

7inch" Programmable colorful touch-screen display

LCD English Display

Microcomputer integrated controller

Functions

Data Record (Max 1500 days x 24hrs/per min), History Curve, USB data download & upload, Remote controlling through PC, Automatic protection alarms, etc.

Program

1~100 programs, 1~100 steps for per program, Max 99hrs 59mins for per program. 99 programs can be interlinked.

Temperature Sensor

Platinum Resistance.PT100Ω

Running Model

Constant/Fix-value Model and Programmable Model

I/O Interface

Communicate port (RS485) to connect test chamber to a PC

Standard Configuration

Test Chamber x 1, Specimen Holder Shelf x 2, φ50mm Cable Access Port x 2 with 2 soft plug, EMO x 1, Tri-color Indicator x 1, etc.

Safety Protection

Power leakage protection/Compressor over-heat, over-load or over-current protection/Dry burning protection.

Voltage

AC220V/380V/440V±10%, 50Hz/60Hz or as per requirement

Ambient Enviromental for application

5°C~+30°C ≤ 85%R.H

Test Standards (not limited to)

IEC60068-2-1

IEC60068-2-2

IEC60068-2-78

IEC60068-2-30


Structure and material:
A. Inner box material: stainless steel plate (SUS # 304 1.0mm thick)
B. Outer box material: stainless steel plate is atomized (SUS # 304 1.0mm thickness)
C. Insulation material: rigid Polyurethane foam and glass wool
D. Supply air circulation system:
(1) 1 90W motor
(2) Stainless steel extension shaft
(3) Multi-wing fan blade (SIRCCO FAN)
E. Box door: single door, single window, open left, handle on right
(1) window 270x360x40mm 3 layers vacuum layer
(2) Plane recessed handle
(3) Back button: SUS # 304


Freezing system:
A. Compressor: Full-density compressor imported from France
B. Refrigerant: Environmentally friendly refrigerant R404A
C. Condenser: Fin-type with cooling motor
D. Evaporator: Fin-type multi-stage automatic load capacity adjustment
E. Other accessories: desiccant, refrigerant flow window, repair valve
F. Expansion system: capacity controlled refrigeration system


Control System:
Originally made by GLAI with full Chinese liquid crystal 7-inch display LCD touch panel, screen talk input data, temperature and humidity can be programmed at the same time, curve display, set value / display value curve. Various alarms can be displayed separately. When a fault occurs, the fault can be displayed on the screen to eliminate the fault and eliminate misoperation. Multiple groups of PID control function, precise monitoring function, and displayed on the screen in the form of data.


1.Controller specifications:
a. Accuracy: temperature ± 0.1 ℃ + 1digit, humidity ± 1% RH + 1digit
b. Resolution: temperature ± 0.1, humidity 1% R.H.
c. Temperature slope: 0.1 ~ 9.9 can be set
d. With upper and lower limit standby and alarm functions
e. Temperature and humidity input signal selection PT 100Ω × 2 (dry bulb and wet bulb)
f. Temperature and humidity conversion output: 4 ~ 20MA
g. 6 group P.I.D control parameter setting, P.I.D automatic calculation, dry and wet ball correction


2.Screen display function:
a. Touch input, display options directly on the screen
b. Temperature and humidity setting (SV) and actual (PV) value are directly displayed (in English)
c. Can display the current execution program number, stage, remaining time and number of cycles
d. Running cumulative time function
e. Temperature and humidity program setting values are displayed in graphical curves, with real-time display of program curve execution functions
f. Separate program editing screen, directly input temperature, humidity and time


3.Program capacity and control function:
a. Available program groups: up to 100 groups
b. Usable memory capacity: 5000SEGMENTS in total
c. Repeatable command execution: each command can reach 9999 times
d. The program is produced in conversation mode, with functions such as editing, clearing, and inserting.
e. SEGMENTS time setting 0 ~ 530Hour59Min
f. With power-off program memory, it can automatically start and continue to execute the program after power is restored (can be stored for more than 10 years)
g. Graphic curve can be displayed in real time during program execution
h. With date and time adjustment function
i. Button and screen lock (LOCK) function


The standard configuration:
A. Power cord 4.0 mm²x3 core
B. One USB export terminal
C. 2 pieces of stainless steel SUS # 304 square punching steel plate rack
D. 2 sets of stainless steel rack rails with adjustable spacing of 40mm
E. 11W energy-saving lighting window light
F. One 50mm test hole on the left side of the machine body, one stainless steel hole cover, one silicone plug.
G. Non-woven standard test piece
H. One set of controller instruction manual and one set of operation manual


Other requirements:
1. Online software
2. Water-cooled cooling system


Surroundings:
(1) Allowable operating temperature range: 0 ~ 35 ℃
(2) Performance guarantee range: 5 ~ 35 ℃
(3) Relative humidity: not more than 85%
(4) Atmospheric pressure: 86 ~ 106Kpa
(5) No strong vibration around
(6) Without direct sunlight or other heat sources
(7) Natural drainage outlet: There must be no bulge in the drain pipe, and the floor of the house must not have extreme incline.

Humidification water supply conditions: pure water
Power supply: 3.5KW 220V 50HZ


Items excluded from valuation:
1. Preparations for the installation site.

2. Electrical equipment on the input side.

3. Electrical wiring works on the input side.

4. Natural drainage piping works.

5. Drain Outlet


http://www.climate-test.com/